EMGA |
ElectroMyoGram Amplifier Module type 691This module is especially designed to amplify EMG signals. A floating input circuit is used to avoid hum and grounding problems. A bargraph display is integrated in the module to visualize the EMG signal. The filters are designed for EMG signals up to 10 kHz. The analog EMG signal is available at a BNC connector on the front panel and also internally on the PLUGSYS system bus. Both outputs can be used for recording on a chart recorder or for data acquisition. A square-wave signal generator with 0.3 mV and 1mV amplitude is incorporated for calibration. For EMG integration an envelope integration circuit is provided. This signal is also available on the front panel and on the PLUGSYS bus. |
Description | |
| Input | floating differential input, isolated barrier internally clamped to 300V, with protection against static discharges and 10 kHz input filter for h.f. suppression |
| Input connector | 5-pin Binder socket with screw lock. The pin connections correspond to those of the standard 3-pin EMG input cable |
| Input impedance | 10 000 MOhm |
| Common mode rejection | 106 dB |
| Filters | Low-pass filter (HIGH CUT OFF) 10 kHz, 3 kHz, 1 kHz, 0.3 kHz
High-pass filter (LOW CUT OFF) 50 Hz, 25 Hz, 10 Hz, 2 Hz |
| Envelope curve | (leakage time constant) 0.2 s, 0.5 s, 1 s |
| Filter indication | by green LEDs on the front panel |
| Gain | GAIN trimmer set to maximum switch x1 x10 x100 in position x0.5 500 5000 50000 in position x1 1000 10000 100000 in position x2 2000 20000 200000 GAIN trimmer set to mimimum switch x1 x10 x100 in position x0.5 100 1000 10000 in position x1 200 2000 20000 in position x2 400 4000 40000 |
| Display | Bargraph 0.75 V/LED |
| Outputs | BNC connector on the front panel for EMG and envelope signal (±10 V), the output signals are also available on the PLUGSYS system bus |
| Calibration | square-wave signal generator with amplitudes of 0.3 mV and 1 mV amplitude |
| Recorder output | The internal outputs are connected to the bus system via jumpers. The EMG signal and the envelope signal are connected to the recorder via a Recorder Output Module ROM. It is also possible to connect a recorder directly to the BNC connectors EMG OUT and ENVELOPE INTEGRATION at the front panel |
| Quick start | for rapid discharge of the coupling capacitors by pressing the QUICK START key |
| Power supply | 5 V/600 mA, through connector from PLUGSYS bus system |
| Dimensions | module for PLUGSYS system, width 8E ( 40.5 mm), height 3U (128.7 mm), depth Europe size (220 mm) |
| Connector | DIN 41612, 96-pin VG |
| Weight | 400 g |
![]()