 EMM

 

ElectroMeter Module type 696 

The HSE Electrometer module is a high-impedance electrometer plug-in amplifier for the PLUGSYS measuring system. It is used to measure concentrations with potentiometric electrochemical sensing electrodes. The main application is recording of pCO2 or Na+, K+ and Ca++ concentrations in biological fluids such as perfusate for isolated perfused organs, using the corresponding electrodes e.g. Type 501, 625 , 601 and 603 (see Chemical Electrodes). The input circuit of the module includes an isolation amplifier (potential separation between sensing electrode and circuit ground of the PLUGSYS measuring system) to avoid measurement errors due to ground loops and leakage currents.

Description 

Input  floating single-ended input, isolated barrier internally clamped to 300V 
Input connector isolated BNC connector 
Input impedance 1015 Ohm 
Input bias current ± 300 fA 
Resolution ± one least significant digit 
Display 3½-digit LED display 
Millivolt Range ± 1200 mV 
Gain x10, x20, x50, x100 set internally using a jumper 
Output On BNC connector on the front panel (±10 V, output impedance 100 Ohm); the output voltage is also available on the PLUGSYS bus. The signal output is assigned to the connecting lines AV-1 to AV-16 through a jumper 
Supression only active on output, adjustable with 10 turn trimmer (maximum ±10 V). The suppression can be switched on and off. 
Simulation a physiological measuring range for calibrating the recording output can be simulated using the 2 corresponding push buttons. For each the simulated value can be adjusted in the range of ±150 mV 
Power supply + 5 Volt / 450 mA (2.25 W) 
Dimensions module for PLUGSYS system, width 8E (40.5 mm), height 3U (128.7 mm),depth Euroboard (220 mm) 
Connector DIN 41612, 96-pin VG 
Weight 0.4 kg 