LLC |
Liquid Level Controler typeThis module is used for monitoring the liquid level in
the oxygenator of the LANGENDORFF Heart apparatus, the WORKING HEART or
ISOLATED LUNG apparatus. A level probe inside the oxygenator controls the
solution |
Description | |
| Probe input | 1/4 bridge, AC voltage signal f = 4kHz through 4-pin connector from the level electrode. |
| lMeasurement principe | comparison of resistance of electrodes immersed in the solution with an internal resistor. |
| Sensitivity | of the level electrode adjustable in the range of 2 kOhm to 50 kOhm |
| Outputs | output sockets for the two motors (pump and sling disc,
adjustable from 2 V to 10 V, 1A max. each) motor on/off signal on the system bus to send this signal to the Flow Measurement module (TTL level) error signal on the system bus to send the status of the LLC to the Flow Measurement (TTL level). |
| Indication | LEVEL HIGH: LED green LEVEL LOW: LED yellow ERROR: LED red PUMP ON: LED green SLING DISC ON: LED green |
| ERROR | to avoid an empty oxygenator when the flow increases too much,
a special circuit (Dry Protection) is installed on this module. If the on-time of the pump is longer than the selected MAX.PUMP ON-TIME an error beep occurs and the pump operates at maximum speed to refill the oxygenator as fast as possible. |
| MAX PUMP ON-TIME | selectable from 1 to 16 s in steps of 1 s (for ERROR detection) |
| TEST | button for rapid filling of the oxygenator (pump operates at max. speed) |
| supply | 5 V/130 mA, +12 V/2 A max, -12 V/1 A max; isolation between level detector and pump control to prevent signal interference from the pump. |
| Dimension | module for 19 inch rack, width 8 E ( 40.5 mm), hight 3 U (128.7 mm), depth Europe-size (220 mm) |
| Connector | DIN 41612, 96 pin. VG |
| Weight | 450 g |
| Accessories | level electrode (suitable for oxygenator) |
![]()
HUGO SACHS ELEKTRONIK-HARVARD
APPARATUS GmbH
Gruenstrasse 1 · D-79232 March-Hugstetten · GERMANY
Phone: +49 (0) 76 65 - 92 00-0 · Fax: +49 (0) 76 65 - 92 00-90