Back

Biochemical Sensors Amplifier modules

 pHMM

pH Measurement Module type 694

The HSE pH Measurement module is a plug-in module for the PLUGSYS measuring system, used to measure pH with pH glass electrodes. The main application is pH recording with the combination pH electrode Type 830 for measurement in biological fluids such as perfusate for isolated perfused organs. The input circuit of the module includes an isolation amplifier (potential separation between sensing electrode and circuit ground of the PLUGSYS measuring system) to avoid measurement errors due to ground loops and leakage currents.

Description 

Input  floating single-ended input, isolated barrier internally clamped to 300V 
Input connector isolated BNC connector 
Input impedance 1015 Ohm 
Input bias current ± 300 fA 
pH Range 0 to 14 
Resolution 0.01 pH 
Display 3½-digit LED display 
Millivolt Range ± 600 mV 
Offset range at pH7 ±100 mV 
Slope 45 mV/pH to 90 mV/pH 
Output 1 Volt per 1 pH on BNC connector on the front panel (±10 V); the output voltage is also available on the PLUGSYS bus 
Suppression 0FF output voltage 0 Volt = pH 0 and 10 Volt = pH 10
0N  output voltage 0 Volt can be adjusted to any pH in the range 0 to 14
 
Slope adjustment 2 point adjustment using calibrated buffer solutions 
Simulation a physiological measuring range for calibrating the recording output can be simulated using the 2 corresponding push buttons. For each the simulated value can be adjusted in the pH range of 0 to 14 
Power supply + 5 Volt /450 mA (2.25 W) 
Dimensions module for PLUGSYS system, width 8E (40.5 mm), height 3U (128.7 mm), depth Euroboard (220 mm) 
Connector DIN 41612, 96-pin VG 
Weight 0.4 kg 


HUGO SACHS ELEKTRONIK-HARVARD APPARATUS GmbH
Gruenstrasse 1 · D-79232 March-Hugstetten · GERMANY
Phone: +49 (0) 76 65 - 92 00-0 · Fax: +49 (0) 76 65 - 92 00-90