Measurement of cardiac output (CO) by the thermodilution method in large animals
![]() |
|
Measurement of cardiac output (CO) by the thermodilution method in rats
|
|
![]() |
The CARDIOMAX III Features / Specifications:
|
|
The new Cardiomax-III thermodilution cardiac
computer interfaces directly to a Pentium desktop or laptop computer
via a very fast RS-232 link. It is intended to be used on a variety
of lab animals from mice to horses. In addition to measuring
cardiac output and associated parameters (SV, HR), the Cardiomax-III
can be used to measure, display, and log ECG (EKG) with its own
low level amplifier and two (2) blood pressures (DP, MP, SP,
dp/dt). Storage of all parameter data is provided in comma separted
ASCII format for easy import into a variety of analysis programs.
The thermo-dilution curve image can also be saved. The Cardiomax-III
can use a variety of thermistor catheters as well as Columbus
Instruments' thermocouple microprobes. The Cardiomax-III's special
design makes it resistant to external radio signal interference
and its small size (35.5 x 29 x 10 cm) makes it very convenient
in a crowded laboratory environment. |
|
![]() |
![]() |
Isolated inputs
(All isolated inputs have over 1500v of low
leakage isolation to the chassis and computer connections.)
Catheter Input
Type: "T" thermocouple or 14k -
7k ohm thermistor
Range: 0 to 50 degrees C.
Injectate Input
Type: "T" thermocouple or 14k -
7k ohm thermistor
Range: 0 to 50 degrees C
ECG Input
Three lead: LA, RA, RL
Range: 50uV to 2mV peak to peak
Blood Pressure Inputs (2)
Range: -100 to + 300 mmHg
Resolution: 1 mmHg,
Accuracy: +/-0.5% full scale, +/-1mmHg
Transducer sensitivity: 5 uV/V/mmHg
Transducer nominal impedance: 300 ohm
Transducer Excitation: 4 volts
Non-isolated connections
High Level ECG Input:
Two ranges: +/-0.25 volt, +/-1.25volt
Foot switch input for starting a C.O. measurement cycle remotely
Injector trigger output for automated C.O. injection
Computer RS-232 serial port connection: 19200 baud
Power requirements
12 watts
120VAC, 60Hz; 240VAC, 50Hz (Factory Selected)
Fuse: 1/2 ampere, type MDA
Dimensions:
Size: 14" x 11.5" x 4" (35.5
x 29 x 10 cm)
Weight: 6 lbs. (2.72 kg)
Ordering Information
TCR Cardiomax III-TCR Thermodilution
Cardiac Output for mice, rats and large subjects
TC Cardiomax III-TC for Thermistor Catheters ONLY for larger animals
only. Not suitable for rodents
R Cardiomax III-R for Micro-Probes ONLY for small animals only.
Requires use of CI Microprobe(s)
Cardiomax III remote foot switch
Cardiomax III replacement Injectate Probe (RTD)
Cardiomax III replacement Injectate Probe (thermocouple)
![]()
HUGO SACHS ELEKTRONIK-HARVARD
APPARATUS GmbH
Gruenstrasse 1 · D-79232 March-Hugstetten · GERMANY
Phone: +49 (0) 76 65 - 92 00-0 · Fax: +49 (0) 76 65 - 92
00-90