Simple Test on Anesthetized Spontaneously Breathing Rodents

  • For measurement of bronchospasmolysis in anesthetized spontaneously breathing rodents (Rat, Guine-Pig, Mouse)
  • Basic simple system
  • Evaluation of typical parameters for respiratory studies (flow, tidal volume, minute volume, respiratory rate, inspiratory and expiratory duration
  • For acute experiments or intubated animals

The rodent is connected to a HSE-PTM or Fleisch pneumotachometer of suitable size through a tracheal or an intubation cannula.When using a Fleisch tube, a special adpater allows the connection of a trachea or intubation canula with a minimum of dead space. A heating transformer for the Fleisch tube is also available to prevent the formation of condensate. The HSE-PTM has especialy been designed to present low dead space and is made of Plexiglas to prevent condensate. The pressure difference produced at the pneumotachometer is converted by a differential pressure transducer VALIDYNE DP 45-14 into an electrical signal proportional to the respiratory flow. The tubing must be kept short to prevent the signal from being distorted by a large dead space between pneumotachometer and differential pressure transducer. A single or multi channel system can be setup using a PLUGSYS system. A pneumotachometer, a differential pressure transducer and a CFBA amplifier are required tor each animal. The HSE- PULMODYN Software is used for data acquisition and evaluation.