This software was especially developed for the investigation
of bronchospasmolytically active substances on conscious animals.
In this experiment the awake animal is placed into a double chamber
plethysmograph box Type 855, and is restricted so that the head
protrudes into the front chamber. The neck is sealed with a soft
diaphragm. The nasal air flow is measured in the front chamber,
the thoracic respiratory flow in the rear chamber. Both measurements
are made with wiremesh measuring screens and differential pressure
transducers. There is a phase shift between the two respiratory
flows from which the PULMODYN PENNOCK software calculates the
specific airway resistance according to PENNOCK. Data acquisition
covers the following signals: pulmonary air flow in the nasal
chamber, pulmonary air flow in the thoracic chamber. From the
flow signal of the thoracic chamber the following parameters are
calculated: respiration rate, tidal volume, maximal inspiratory
flow, maximal expiratory flow. From the phase shift between nasal
and thoracic flow phase shift and specific airway resistance are
calculated. During data acquisition all acquired signals and derived
parameters can be displayed on the screen.
The system covers eight channels. This makes it possible to experiment
simultaneously on 4 plethysmograph boxes. All 8 signals are sampled
at 1 kHz. This means that each channel is measured 1000 times
per second, or measurements on each channel are made once every
millisecond. This permits measuring phase shifts of some milliseconds
and calculate from this delay the specific airway resistance.