Back

Transducers for respiratory mechanics measurements

PTM type 378/1.2 for rats

PTM type 378/0.9 for mice 

HSE - PNEUMOTACHOMETER PTM type 378/X for rats and mice 

The HSE pneumotachometer (PTM) type 378/X is a transducer for airflow measurement. It is especially designed for respiratory flow measurement on isolated lung of rats, guinea pigs or mice. In combination with the silicon rubber tubing and the tracheal cannula the PTM can also be used for acute experiments (tracheotomy). The main feature of the new design is the small dead space. The PTM is made of transparent perspex, this allows a good view of the inside during the experiment or the cleaning. If during the experiment mucus comes up into the PTM it’s easy to detect and prevent for measurement errors caused by artificial increasing of the airway resistance due to PTM obstruction. In addition the perspex has the advantage over metall to present low temperature conductivity so that no heating is needed to prevent from condensation of the moisture of the expired air. The flow resistance consists only of straight boring and is therefore easy to clean.
The basic principle of the transducer is the generation of a pressure proportional to the airflow across a flow resistance. For the measurement of the pressure we recommend the VALIDYNE differential pressure transducer type DP 45-14.

type 378/1.2 for rats type 378/0.9 for mice
Nominal sensitivity: 10 mmH2O for 27 ml/s 10 mmH2O for 10.5 ml/s
Nominal flow range: ±27 ml/s ±10.5 ml/s
Flow resistance, approx: 0.4 mmH2O/ml/s 1.0 mmH2O/ml/s
Dead space, approx.: 50 µl 25 µl


 

 

differential pressure transducer DLP2.5

  • Suitable for airflow measurements using pneumotachs
  • For very low differential pressure measurements
  • Range ±2.5 cm H20
  • Simple rod mounting

SPECIFICATION
Range: ±2.5 cm H20 (PTM and FLEISCH tube) 
Overpressure: 

250 cm H20

Natural frequency:  greater than 50 Hz 
Measuring system:  Full bridge
Pressure cavity volume:   120 mm3 
Excitation voltage:  5 VDC 
Thermal zero shift:  ±1.5%FS (-25 to 85°C) 
Sensitivity: 

65 to 67 mV/cm H20 at 5 V excitation

Linearity:  ±0.25% FS 
Inlet Nozzle, ID x OD x L:  1,8 x 3,5 x 8.0 mm 
Housing Size, H x W x D:  24 x 42 x 36 mm (0.9 x 1.7 x 1.4 in) 
Weight:  190 g (6.7 oz) 

 

 

differential pressure transducers VALIDYNE

Type DP 45-14

  • for connection to pneumotachometer FLEISCH or HSE - PTM or plethysmograph box
  • range ± 2 cm H2O
  • inductive system, needs CF-bridge (or CFAB with DC-bridge amplifier)

Type DP 45-24

  • for measuring oesophageal pressure
  • range ± 22.5 cm H2O
  • inductive system, needs CF-bridge (or CFAB with DC-bridge amplifier)

SPECIFICATION for both:
Ranges: DP 45-14: 2 cm H2O (FLEISCH tube)
DP 45-24: ± 22.5 cm H2O (oesophageal pressure)
 
Volumetric displacement:  0.006 mm3 FS 
Natural frequency:  greater than 600 Hz 
Measuring system:  half bridge inductance (2 × 20 mH nominal) 
Pressure cavity volume:   6.5 mm3 
Excitation voltage:  5 V at 5 kHz 
Thermal zero shift:  0.2% /10oK (25°C ± 40°K) 
Sensitivity:  25 mV/V FS nominal 
Accuracy:  ± 0.25% FS 
Size:  diam.: 57 mm, width: 20.5 mm 
Weight:  0.6 kg 

 

Differential pressure transducer MPX Type 399/2

for measuring oesophageal pressure

specification:
Pressure range: ±100 cm H2O (±100 mbar) 
Sensitivity:  0.3 . . . 0.8 mV/mbar (excitation: 5 V) 
Linearity:  ±1.5% 
Thermal zero shift:  5 mbar (0–85°C) 
Input resistance:  400 . . . 550 Ohm
Output resistance:  600 . . . 1000 Ohm
Offset voltage:  1 mV max. 
Excitation voltage:  0 . . . 5 Volts DC or AC 
Overpressure:  ±1000 mbar (750 mmHg) 
Sizes:  housing: 42 mm × 36 mm × 24 mm 
weight:  190 gr. 
Inlet/outlet nozzle:  i.d. 2.0 mm; o.d. 4.7 mm; length 9.0 mm 
application:  only for dry air 

 


HUGO SACHS ELEKTRONIK-HARVARD APPARATUS GmbH