Transducers for tissues contractions measurements

Isometric transducer

  Force transducer for recording tissue contractions under strict isometric conditions. This transducer can be used for smooth muscle (rings, stips...), for cardiac muscle (atrium, papillary muscle, purkinje fiber...) and for squelettal muscle (diaphragm...). The transducer exists in two versions depending on the sensitivity: F30 and F10. These transducers can be connected to any strain gauge bridge amplifier (DBA, TAM-A, TAM-D...)
specification
 F30  F10
Force range:  ±0.3 N (±30 grams)  ± 0.1 N (± 10 grams)  
Linear range:  ±80 cN  ±20 cN 
Maximum load:  ±3 N  ± 0.5 N 
Displacement range:  ±0.03 mm  ± 60 µm 
Sensitivity:  0.35 mV/V · g  2 mV/V · g 
Isometric Quotient IQ:  1 µm/g  6 µm/g 
Excitation AC or DC:  5 V (max. 10V) 5 V (max. 10 V) 
Full-scale output (nominal):  ±10 mV/V  ± 20 mV/V 
Seismic sensitivity:  15 mN/acceleration of gravity   15 mN/acceleration of gravity 
Natural frequency (nominal):  440 Hz 180 Hz 
Bridge resistance (nominal):  1 kOhm - full resistance bridge, to be connected to any bridge amplifier   1 kOhm – full ohmic bridge, to be connected to any bridge amplifier 
Hysteresis:  < 0.1 %  
Non-linearity:  <0.1 %   
Zero drift:  < 10 mg/°C at 20 to 40°C 
< 0.03 % FS/°C 
 
Shift at 10 g load:  < 1 % after 10 minutes   
Resolution (noise):  2 mg  0.5 mg 
Maximum load:  300 %  100 % 
Breaking load:  1000 %  300 % 
Sensitivity to transverse load:   < 0.1 %   
Maximum allowed transverse load:  < 300 %   

K30 Force transducer for recording tissue contractions under auxotonic conditions. This transducer can be used for smooth muscle (rings, stips...), for cardiac muscle (atrium) and especially for squelettal muscle (diaphragm...). The transducer exists in four versions depending on the sensitivity: K30, K100, K300 and K1000. These transducers can be connected to any strain gauge bridge amplifier (DBA, TAM-A, TAM-D...) 
specification
K30  K100 K300  K1000 
Force range: ± 30 g-force  
± 0.3 N
± 100 g-force 
± 1N
± 300 g-force
± 3 N
± 1000 g-force
± 10 N
Displacement:  ± 2.5 mm  ± 2 mm  ± 3 mm   ± 1 mm 
Sensitivity (force):   100 µV/g-force/V  30 µV/g-force/V  10 µV/g-force/V  3 µV/g-force/V 
Sensitivity (displacement):  1200 µV/mm/V  0.7 mV/mm/V  0.8 mV/mm/V   1.2 mV/mm/V 
Natural frequency without load:  110 Hz  170 Hz  200 Hz   400 Hz 
Range with additional spring:   ± 200 g, ± 2.5 mm  - / -  - / -   - / -  
Sensitivity with additional spring:  10 µV/g/V  - / -  - / -   - / -  
Natural frequency with additional
spring:
 
230 – 270 Hz  - / -  - / -   - / - 
Resolution:  5 mg  15 mg   50 mg   150 mg 
Input impedance (pins 1–5): 350 Ohm 
Output impedance (pins 2–4): 350 Ohm 
Supply voltage: max. 10 V  

Isotonic transducer

Lever transducer B40. This displacement transducer is used for measurement of isotonic contractions of smooth muscles (ileum, bladder, uretra....) This transducer can be connected to any strain gauge bridge amplifier (DBA, TAM-A, TAM-D...)

 

specification
Range: ±20 mm (0 – 40 mm)
Linearity:   ±2 % full scale
Zero drift:  <0.03%/°C FS
Frequency range: 0 – 10 Hz
Full bridge:   4 x 300 Ohm
Frictional torque of lever:  0.1 mN · cm
Output voltage:  ±10 mV, adjustable up to maximum ±40 mV, internal resistance of 300 Ohm.
The maximum output voltage represents the full lever deflection and the full measuring range.
Power supply:  5 V DC or AC (maximum 8 V) 

 