harvardapparatus.com is a data controller for the purposes of the General Data Protection Regulation ("GDPR"). By proceeding to tick the consent box at right, you consent to harvardbioscience.com processing your personal data, under the GDPR and any other applicable legislation, that we collect from you in accordance with our Privacy Policy. If you wish to withdraw your consent for us to process your personal data please contact our Compliance Team. Please address any questions, comments and requests (including your right to withdraw your consent to process your personal data) regarding our data processing practices to our Compliance Team.
Allow

Extracorporeal In-Line Flow Probes for TTFM-2 Flowmeter Modules

Price on Request
SKU
INLINEFP-GRP

Extracorporeal in-line flow probes monitor perfusion flow on isolated organs and are built into the perfusion lines (tubing). For use with PLUGSYS Transit Time Flowmeter Module (TTFM-2).

These extracorporeal in-line probes are available in sizes from 1.0 to 8.0 mm. All in-line flow probes are cannulated and options are available for a larger range of species.

Grouped product items
Item No. Description Qty
73-4753 Inline Flow Probe, type 1N for TTFM-2. cannulated, 1 mm ID Please login to add to Quote
73-4754 Inline Flow Probe, type 2N for TTFM-2, cannulated, 2 mm ID Please login to add to Quote
73-4755 Inline Flow Probe, type 4N for TTFM-2, cannulated, 4 mm ID Please login to add to Quote
73-4947 Inline Flow Probe, type 8N for TTFM-2. cannulated, 8 mm ID Please login to add to Quote
73-5131 Inline Flow Probe, type 12N for TTFM-2. Cannulated, 12 mm ID Please login to add to Quote
73-5233 Inline Flow Probe, type 16N for TTFM-2. Cannulated, 16 mm ID Please login to add to Quote

Extracorporeal in-line flow probes monitor perfusion flow on isolated organs. For use with PLUGSYS Transit Time Flowmeter Module (TTFM-2) in isolated organ systems such as abdominal organs or lungs, or for any application where a flow in a tube must be measured.

These probes are embedded into a perspex holder which has two ports. There are installed into an in-line tubing. These probes are available in sizes from 1.0 to 8.0 mm. All in-line flow probes are cannulated, and options are available for a range of species.

    Resolution Low Flow Normal Max Flow Zero Offset Absolute Accuracy Relative accuracy
Part No
Model ml/min 1V output in ml/min 1V output in ml/min 5V output in ml/min ml/min % of reading %

73-4753

1N

0.05

5

20

100

±0.2

±10

±2

73-4754

2N

0.1

25

100

500

±1

±10

±2

73-4755

4N

0.8

100

400

2L

±4

±10

±2

73-4946

6N

2.0

250

1L

5L

±10

±10

±2

73-4947

8N

4.0

500

2L

10L

±20

±10

±2

73-5131

12N

8.0

1L

4L

20L

±40

±10

±2

73-5233

16N

20

2.5L

10L

50L

±100

±10

±2

Search engine powered by ElasticSuite