Extracorporeal In-Line Flow Probes for TTFM-2 Flowmeter Modules
Extracorporeal in-line flow probes monitor perfusion flow on isolated organs and are built into the perfusion lines (tubing). For use with PLUGSYS Transit Time Flowmeter Module (TTFM-2).
These extracorporeal in-line probes are available in sizes from 1.0 to 8.0 mm. All in-line flow probes are cannulated and options are available for a larger range of species.
Extracorporeal in-line flow probes monitor perfusion flow on isolated organs. For use with PLUGSYS Transit Time Flowmeter Module (TTFM-2) in isolated organ systems such as abdominal organs or lungs, or for any application where a flow in a tube must be measured.
These probes are embedded into a perspex holder which has two ports. There are installed into an in-line tubing. These probes are available in sizes from 1.0 to 8.0 mm. All in-line flow probes are cannulated, and options are available for a range of species.
Resolution | Low Flow | Normal | Max Flow | Zero Offset | Absolute Accuracy | Relative accuracy | ||
Part No |
Model | ml/min | 1V output in ml/min | 1V output in ml/min | 5V output in ml/min | ml/min | % of reading | % |
73-4753 |
1N |
0.05 |
5 |
20 |
100 |
±0.2 |
±10 |
±2 |
73-4754 |
2N |
0.1 |
25 |
100 |
500 |
±1 |
±10 |
±2 |
73-4755 |
4N |
0.8 |
100 |
400 |
2L |
±4 |
±10 |
±2 |
73-4946 |
6N |
2.0 |
250 |
1L |
5L |
±10 |
±10 |
±2 |
73-4947 |
8N |
4.0 |
500 |
2L |
10L |
±20 |
±10 |
±2 |
73-5131 |
12N |
8.0 |
1L |
4L |
20L |
±40 |
±10 |
±2 |
73-5233 |
16N |
20 |
2.5L |
10L |
50L |
±100 |
±10 |
±2 |