pH Amplifier for PowerLab
The pH Amplifier is a dual front-end amplifier for PowerLab suitable for use with pH, ion selective and potentiometric redox electrodes. It can be used as a basic pH meter, or in conjunction with the in-built temperature amplifier to provide full temperature compensation.
The FE165 pH Amplfier is a dual front-end suitable for use with pH, ion selective and potentiometric redox electrodes. It can be used as a basic pH meter, or in conjunction with the in-built temperature amplifier to provide full temperature compensation
With a suitable electrode, parameters such as pH, redox potential, dissolved carbon dioxide and ammonia can be measured in aqueous solution. Ion selective electrodes are available for many ionic species including sodium, potassium, calcium, chloride, fluoride, nitrate and sulfide.
Compatibility:
The pH Amp is compatible with all current PowerLab and MacLab models.
The Following are Some of the Transducers and Accessories Suitable for Use With the pH Amp:
• MLA060 Redox Electrode
• MLT5733 pH Electrode - Tuff Tip for Student Use
• MLA042 pH Electrode
The ML165 pH Amp is Supplied Completed With the Following:
• I²C Cable (9-Pin to 9-Pin)
• BNC to BNC Cables (2)
• MLT404/A RTD Temperature Probe
• pH Amp Owner's Guide
Specifications | 77-0291 | 77-0052 | 77-0051 | 77-0093 |
---|---|---|---|---|
Certifications | CE | |||
Control Port | I²C Provides control and power | |||
DC Drift | 3 µV/°C | |||
Dimensions H x W x D English | 2.2 x 4.7 x 10.2 in | |||
Dimensions H x W x D Metric | 55 x 120 x 260 mm | |||
Enclosure Size L x W x H English | 9.8 x 7.9 x 2.6 x in | |||
Enclosure Size L x W x H Metric | 250 x 200 x 65 x mm | |||
Frequency Response | DC ? 25 kHz nominal | |||
Gain | x 1 (internal differential buffering) | |||
Input Configuration | Differential (fixed) | |||
Input Connection Type | 8-Pin DIN | |||
Input Impedance | ~1 MΩ | |||
Maximum Input Voltage | ±10 V | |||
Maximum Number of Pods | 4 pods | |||
Maximum Power Needs | ~1 W | |||
Model | FE165 | MLA060 | MLA042 | MLT5733 |
Model | ML305 | |||
Operating Conditions | 0 ? 35 °C, 0 ? 90 % humidity (non-condensing) | 0 to 60 °C | -5 to 100°C | |
Operating Conditions | 0 to 35 °C, 0 to 90 % Humidity (non-condensing) | |||
Output Voltage | Same as Pod output voltage (±5 V max) | |||
Package | 1 | 1 | 1 | 1 |
Power | Pod: ±5 V, 100 mA maximum per supply rail | |||
Product Family | Data Acquisition | Data Acquisition | Data Acquisition | Data Acquisition |
Temperature Amp Amplifier Output | 0 V at 0 °C, 50 mV/ °C (factory set) | |||
Temperature Amp Input Connection | 3 pin mini-audio jack | |||
Temperature Amp Output Ranges | 500µV to 10V | |||
Temperature Amp Temp Accuracy | ±0.2 °C | |||
Temperature Amp Temp Probe Type | RTD 100 Ω Type (supplied with unit) | |||
Temperature Amp Temperature Range | ±100 °C | |||
Weight English | ~3 lb | |||
Weight English | 2.6 lb | |||
Weight Metric | 1.2 kg | |||
Weight Metric | 1.5 kg | |||
pH Amplification Range | +/-2V down to +/-200 µV full scale in 13 steps | |||
pH Amplifier Amplifier Noise | <1 µVrms (<4 |