harvardapparatus.com is a data controller for the purposes of the General Data Protection Regulation ("GDPR"). By proceeding to tick the consent box at right, you consent to harvardbioscience.com processing your personal data, under the GDPR and any other applicable legislation, that we collect from you in accordance with our Privacy Policy. If you wish to withdraw your consent for us to process your personal data please contact our Compliance Team. Please address any questions, comments and requests (including your right to withdraw your consent to process your personal data) regarding our data processing practices to our Compliance Team.
Allow

PLUGSYS External Input Module (EIM)

SKU
GRP-EIM

For linking external equipment to the PLUGSYS system for connection to HA-HSE data acquisition hardware

Grouped product items
Item No. Description Qty
No options of this product are available.

The External Input Module (EIM) is used for feeding external analog signals into the PLUGSYS Measuring System, such as a flow signal from an external blood flowmeter, a force signal from an external bridge amplifier or distance signals measured using a sonomicrometer. 

The EIM module (73-0222) consists of the two parts: EIM-A (Adjust) and EIM-B (BNC). The modules EIM-A and EIM-B normally fit in the PLUGSYS Mainframe from the front. With an optional board it is possible to install the module EIM-B on the rear to have more space on the front.

Specifications73-0222
Accessories2 BNC-BNC input cable, operating instructions
ConnectorDIN 41612, 96-pin VG
Dimensions (H x W x D) EnglishEIM-A 5.1 x 1.6 x 8.7 in
EIM-B 5.1 x 1.6 x 2 in
Suppression 5.1 x 1.6 x 4 in
Dimensions (H x W x D) MetricEIM-A 128.7 x 20.2 x 220 mm
EIM-B 128.7 x 20.2 x 50 mm
Suppression 128.7 x 20.2 x 100 mm
GainFrom 0.25 to 100 adjustable by 10-turn trimmer together with range switch x0.5/x1/x2, corresponding to max. input voltage range of ±0.1 to ± 10 V
Input4 analog inputs through BNC sockets on module EIM-B
OutputThrough links on circuit board (jumpers) signal outputs are fed directly into bus distribution board of PLUGSYS measuring system
PLUGSYS Width*3 slot units
Power Supply5 V, 300 mA from PLUGSYS measuring system
Weight0.6 lb (0.25 kg)
Search engine powered by ElasticSuite