Compact Transducer Amplifier
Compact Transducer Amplifier
The Compact Transducer Amplifier (CTA) is a DC bridge-based transducer amplifier and signal conditioner for HSE PLUGSYS Transducers. It can be used with pressure, force, displacement or other bridge-based transducers. See the Item List and Details Section for the compatible list of transducers. It is factory adapted and calibrated for use with a specific transducer model. Once configured it cannot be modified for use with any other transducer. The CTA and transducer must be ordered together. Please see our Multi-Sensor Amplifier if you intend to use the amplifier for multiple transducer types or you require a digital display of average values.
The Compact Transducer Amplifier (CTA) is a DC bridge-based transducer amplifier and signal conditioner for HSE PLUGSYS Transducers. It can be used with pressure, force, displacement or other bridge-based transducers. See the Item List and Details Section for the compatible list of transducers. It is factory adapted and calibrated for use with a specific transducer model. Once configured it cannot be modified for use with any other transducer. The CTA and transducer must be ordered together. Please see our Multi-Sensor Amplifier if you intend to use the amplifier for multiple transducer types or you require a digital display of average values.
It comes with a power supply suitable for all countries (90 to 264 V). The CTA powers the selected transducer (excitation voltage 5 V), amplifies the signal and outputs a signal voltage (range ±4 V). Trimmer adjustments are available for gain (30 to 100%) and zero offset.
The CTA analog output signal can be input to a suitable data acquisition system or other device (e.g. the Harvard Apparatus PHD Ultra Constant Pressure syringe pump) that can accept a high-level signal in the range of ±4 V.
The CTA includes: enclosed amplifier, power supply with country either a EU or US mains cable (power cord)*, HSE transducer adapter cable, one BNC-BNC output cable (length 2 m), operating instructions.
Compatible Transducers:
ItemNo | Description |
73-3862 | PRESSURE TRANSDUCER APT300 FOR HSE PLUGSYS AMPLIFIERS, RANGE -300 TO 300mmHg WITH 6 PIN HSE CONNECTOR (73-2909) CABLE LENGTH 3m (10ft)A |
73-0020 | LOW RANGE BLOOD PRESSURE TRANSDUCER P75 FOR HSE PLUGSYS MODULE A |
73-4479 | Manual Pressure Calibrator, Range 0-300 mmHgA |
73-3882 | Differential Low Pressure Transducer DLP2.5, range +- 2.5 cmH2O, HSE Connector |
73-0064 | Differential Pressure Transducer MPX, Range +- 100 cmH2O, HSE ConnectorA |
73-4986 | FT20 ISOMETRIC FORCE TRANSDUCER +/- 20g FOR HSE AMPLIFIER CABLE INCLUDED |
73-4987 | FT50 ISOMETRIC FORCE TRANSDUCER +/- 50g FOR HSE AMPLIFIER CABLE INCLUDED |
Please see our complete line of Digital Pressure Calibrators.
Data Acquisition Hardware and Software
Please see our complete line of Data Acquisition Hardware and Software Solutions for Physiologic signals if you need to monitor and record the signal output.
Frequency Response | DC - 1500 Hz |
Transducer Excitation Voltage | 5 VDC, 100 mA max , Ri = 10 Ω |
Output Voltage Range | ±4 V |
Gain | 1 to 1000, factory adapted to the transducer Gain variable 30 to 100% fine adjustment with trimmer |
Zero Offset | Manually adjustable with trimmer |
Power Supply | External power supply Input voltage 100 to 240 VAC, 50/60 Hz Output voltage 12 VDC/300mA |
Power Consumption | 100 mA max. |